EBC - ON DEMAND
European Bifurcation Club 2015 – EBC 2015 – Athens, Greece
Novel insights from bench and simulation
Shear stress analyses of absorb and mirage brs
Author: Patrick W. Serruys, MD, PhD, Erasmus Medical Center, Rotterdam (NL)
CONCLUSION
As expected from in silico studies, we found higher ESS in the Mirage BRMS which has thinner ovoid struts and helical coil design compared to the Absorb BVS. Helical coil designed thinner ovoid-shaped struts appear to cause less flow disturbances than the design of zig-zag hoops with thick rectangular-shaped struts.