Safety of hydrophilic polymer-coated guide wires used for side-branch protection during bifurcation stenting and POT
Author: Massoud Leesar, MD Professor of Medicine Section Chief, Interventional Cardiology, Baker Dean Endowed professor University of Alabama, Birmingham
CONCLUSION
- There was no wire fracture by jailing polymer- coated wires
- The extent of polymer shearing was significantly higher with Whisper wire than Runthrough wire probably as a result of complete polymer coverage
- The used of polymer-coated guide wires was associated with embolization of polymer to the myocardial microvasculature system and was an independent predictor of PMI
- The impact of polymer shearing on myocardial infraction warrants future studies