Further Optimization of Kissing Balloon Inflation: Insight from computer simulations

European Bifurcation Club 2014 - EBC 2014 - Bordeaux, France Bifurcation stenting: learning from bench and simulation Further Optimization of Kissing Balloon Inflation: Insight from computer simulations Author: Peter Mortier, MD, FEops / Ghent University, Belgium CONCLUSION Side branch access: A good kissing inflation does not guarantee optimal side branch access - A large variability was observed with each stent design Inflating the side branch balloon first optimizes side branch access Stent strut apposition: Malapposed struts are always present near the side branch ostium Reducing the pressure in the SB balloon does not increase the amount of malapposed stent struts The elliptical deformation in the proximal segment can be reduced by reducing the pressure in the side branch balloon prior to inflating the main branch balloon.

Bifurcation stenting simulation: the vessel

European Bifurcation Club 2010 - EBC 2010 - Budapest, Hungary From bench to simulation Bifurcation stenting simulation: the vessel Author: Peter Mortier - IBiTech - bioMMeda Ghent University, Belgium, 2 FEops, Gent, Belgium

John Doe provisional left main stenting