Is respect of bifurcation branching laws the best option for LM stenting?

EBC - ON DEMAND

Is respect of bifurcation branching laws the best option for LM stenting?

European Bifurcation Club 2011 – EBC 2011 – Lisbon, Portugal

Mr John Doe left main stenting

Is respect of bifurcation branching laws the best option for LM stenting?

Author: Ghassan S. Kassab, Thomas Linnemeier Chair Professor Biomedical Engineering, Indiana/Purdue University

 

CONCLUSION

  • Since Dm7/3 = Dl7/3 + Ds 7/3 is derived from flow.structure scaling law and conservation of mass, DCAD will not affect the bifurcation law. Hence, despite DCAD, if the diameter of the two ‘normal’ vessel segments is known, the third diseased segment can be determined from the bifurcation law
  • The coefficient of structure-structure scaling laws can be used to diagnose DCAD

 